Which of the following are true regarding large-volume jet nebulizers?

a. A variable air-entrainment port allows air mixing and different FiO2 values.
b. Liquid particles are generated by mechanical vibration energy.
c. They are pneumatically powered and attach directly to a flowmeter.
d. Baffling causes impaction or removal of large, unstable particles.
a. 1 and 3 only
b. 1, 2, and 3 only
c. 2 and 4 only
d. 1, 3, and 4 only


ANS: D
As depicted in Figure 38-13, these devices are pneumatically powered, attaching directly to a flowmeter and compressed gas source. Liquid particle aerosols are generated by passing gas at a high velocity through a small "jet" orifice. The resulting low pressure at the jet draws fluid from the reservoir up to the top of a siphon tube, where it is sheared off and shattered into liquid par-ticles. The large, unstable particles fall out of suspension or impact on the internal surfaces of the device, including the fluid surface (baffling). The remaining small particles leave the nebulizer through the outlet port, carried in the gas stream. A variable air-entrainment port allows air mix-ing to increase flow rates and to alter FiO2 levels.

Health Professions

You might also like to view...

Ask 10 people their height. Make a list and arrange it in order by height

What will be an ideal response?

Health Professions

Which of the following organizations has made recommendations regarding verbal prescriptions?

MERP ISMP MEDMARX MedWatch

Health Professions

Where is the bacterium that causes plague most commonly found?

a. on rodents b. in human feces c. in damp, moist, dark areas d. in decomposing vegetation

Health Professions

An intact female rabbit is a/an_____

Fill in the blank(s) with correct word

Health Professions