Which of the following are essential components of a continuous positive airway pressure (CPAP) flow system?

1. Blended source of pressurized gas
2. Nonrebreathing circuit with reservoir bag
3. Low-pressure or disconnect alarm
4. Expiratory threshold resistor
a. 3 and 4 only
b. 1, 2, and 4 only
c. 1 and 4 only
d. 1, 2, 3, and 4


ANS: D
A breathing gas mixture from an oxygen blender flows continuously through a humidifier into the inspiratory limb of a breathing circuit. A reservoir bag provides reserve volume if the patient's inspiratory flow exceeds that of the system. The patient breathes in and out through a simple valveless T-piece connector. A pressure alarm system with manometer monitors the CPAP pres-sure at the patient's airway. The alarm system can warn of either low (usually due to a disconnec-tion) or high system pressure. The expiratory limb of the circuit is connected to a threshold resis-tor, in this case a water column (H).

Health Professions

You might also like to view...

The sense of presence of the massage therapist's hands on the client's body during massage should feel:

A) Deliberate and nervous B) Confident and warm C) Hesitant and fearful D) Full and unsure

Health Professions

________ in peripheral tissues remove cellular debris and respond to invasion by foreign compounds or pathogens

A) Lymphocytes B) Immunocytes C) Andocytes D) Phagocytes

Health Professions

How many categories of the ARFF are defined by the NFPA?

a. Two b. Three c. Four d. Five

Health Professions

A localized bacterial infection, causing the footpads to become inflamed and develop sores is referred to as _________________ foot

Fill in the blank(s) with correct word

Health Professions