What is the breath termination for a CareFusion V.I.P. Bird Infant-Pediatric Ventilator set to de-liver pressure support at tidal volume = 250 mL and peak expiratory flow 15 L/min?

a. Time cycle at 1 second c. Flow cycle at 1 L/min
b. Time cycle at 1.25 second d. Flow cycle at 3.75 L/min


ANS: D
Breaths in the volume-cycled mode on the CareFusion V.I.P. Bird Infant-Pediatric Ventilator are terminated using flow-cycling. For delivered tidal volume greater than 200 mL, the flow-cycle value is 25% of peak flow.