The therapist is monitoring the flow, volume, and pressure scalars of a patient receiving mechanical ventilation. After the first exhalation, the therapist initiates an expiratory pause. Based on the pressure-time tracing in particular, what will the therapist observe?

A. plateau pressure
B. auto-PEEP
C. applied PEEP
D. applied and auto-PEEP


?ANS: D
A. Incorrect response: See explanation D.
B. Incorrect response: See explanation D.
C. Incorrect response: See explanation D.
D. Correct response:
By performing an expiratory pause (EP), the therapist will be able to measure the intrinsic PEEP. The applied PEEP will be observed and measured with every breath without the therapist having to conduct any special maneuvers. The difference between the measurement obtained as a result of the expiratory pause and the applied PEEP will equal the auto-PEEP.

Health Professions

You might also like to view...

What is the normal value for ETCO2 in a healthy young adult?

A. 5% B. 8% C. 15% D. 18%

Health Professions

All providers are required by HIPAA to use:

a. the same code sets, identifiers, and transactions when healthcare information is transmitted b. the Internet c. the extranet d. encryption

Health Professions

The only difference in interpretation criteria between sinus rhythm, sinus bradycardia, and sinus tachycardia is the heart rate

Indicate whether the statement is true or false

Health Professions

Which of the following laws requires most legend drugs to be packaged in child-resistant containers?

Harrison Narcotics Tax Act Poison Prevention Packaging Act Food, Drug, and Cosmetic Act Drug Listing Act

Health Professions