Which of the following statements is true concerning the operation of a transcutaneous carbon dioxide electrode?
a. PtcCO2 readings are slightly lower than the PaCO2 value.
b. Electrodes should be calibrated when they are repositioned.
c. Heating the transcutaneous PCO2 (PtcCO2) to 43°C will adversely affect its operation.
d. Before its use, the electrode should be calibrated to 20% and 40% CO2.
ANS: B
Manufacturers typically suggest calibration of an electrode each time it is repositioned. PtcCO2 readings are slightly higher than the PaCO2 value because of the higher metabolic rate at the site of the electrode caused by heating the skin. The PtcCO2 electrode must be heated to 42° C to 44° C. The PtcCO2 electrodes are calibrated with a two-point calibration procedure by using 5% CO2 and 10% calibration gases.
You might also like to view...
When there are too many new, or immature, cells that are not functional then ____________________ has occurred.
Fill in the blank(s) with the appropriate word(s).
Intractable, chronic, posttraumatic headache is reported with code _____.
Fill in the blank(s) with the appropriate word(s).
Interventions change behavioral precepts, which are the second level of the disease prevention model.?
Indicate whether the statement is true or false
What form of nuclear imaging emits positrons, causing gamma rays to be discharged?
a. MRI b. PET c. Ultrasound d. Fluoroscopy