Absorbance or fluorescent sensors are used with which of the following?
a. Capnography c. In vitro blood gas analysis
b. Photoplethysmog-raphy d. In vivo blood gas analysis
ANS: D
In vivo blood gas monitors use optical sensors that are generally categorized in terms of how they modify the initial optical signal; they are classified as either absorbance or fluorescent sensors. Absorbance sensors absorb a fraction of the incident light as it is transmitted down the fiber and through the microcuvette. The amount of light transmitted is proportional to the concentration of the analyte in question. Fluorescent sensors use dyes that fluoresce when they are struck by light in the UV or near-UV visible range. Light from the monitor is transmitted to the microcuvette containing the dye. The concentration of the analyte in question can be measured by determining the ratio of fluorescent light emitted to the original excitation light signal.
You might also like to view...
Which of the following are signs or symptoms of a CVA?
1) headache 2) paralysis 3) nausea 4) dizziness a. 1, 2 b. 1, 2, 3 c. 2, 4 d. 1, 2, 3, 4
Describe three (3) reasons that meeting OSHA standards for safety is beneficial to the healthcare institution
What will be an ideal response?
Word Scramble: Sugar Gliders
SPA
In angiography, contrast medium is injected into a(n) ________ by way of a catheter and threaded through the vessel until it reaches the correct site.
A. artery B. vein C. either artery or vein D. neither artery nor vein