Which sequence of steps for plate fixation is CORRECT?
a. Drill all holes, place plate, tap, measure with depth gauge, and insert screws.
b. Drill all holes, place plate, measure with depth gauge, tap, and insert screws.
c. Place plate, drill one hole at a time, tap, measure with depth gauge, insert screw, and repeat steps for remaining holes.
d. Place plate, drill one hole at a time, measure with depth gauge, tap, insert screw, and repeat steps for remaining holes.
d