Which of the following statements concerning the use of AutoFlow for neonates on the Dräger E-4 is true?
a. AutoFlow is an adult mode and should not be used with neonates.
b. The ventilator's internal flow sensor is al-ways used with neonates on AutoFlow.
c. Two test breaths determine the appropriate pressure to deliver volume in AutoFlow.
d. In AutoFlow, the highest pressure to de-liver the volume is the high-pressure limit setting.
ANS: C
When the flow sensor is attached during AutoFlow in neonates, the unit provides two test breaths, first with a pressure of 5 cm H2O and then with a pressure equal to 75% of that required to deliver the set volume.