What are some key causes of patient-ventilator asynchrony and increased work of breathing dur-ing pressure-triggered volume-controlled continuous mandatory ventilation?

1. Improper trigger setting
2. Insufficient inspiratory flow
3. High peak airway pressures
a. 1 and 2 only
b. 1 and 3 only
c. 2 and 3 only
d. 1, 2, and 3


ANS: A
If sensitivity is set too low, such that considerable effort is necessary to trigger the ventilator, pa-tient-ventilator asynchrony occurs. A pressure sensitivity of −0.5 to −1.5 cm H2O or flow sensi-tivity of 1 to 2 L/min is regarded as optimal. Inspiratory flow must be set to meet the patient's inspiratory demand. An insufficient inspiratory flow can cause patient-ventilator asynchrony and increased work of breathing.

Health Professions

You might also like to view...

Antibiotics are used to control the viruses.

Answer the following statement true (T) or false (F)

Health Professions

A career passport or portfolio is a professional way to highlight your knowledge, abilities, and skills as you prepare for employment or extended education

Indicate whether the statement is true or false

Health Professions

Where does the victim's arm go in the one-person walking assist?

a. across the victim's chest b. around the rescuer's neck c. linked in the rescuer's arm d. around the rescuer's waist

Health Professions

Which of the following is not a type of digital image receptor?

A) Photostimulable phosphor (PSP) plate B) Scanner C) Complementary metal oxide semiconductor active pixel sensor (CMOS-APS) D) Charge-coupled device

Health Professions