Which of the following statements concerning the CareFusion 3100A and 3100B is true?

a. There is a mean pressure limit control on both models.
b. Piston centering is automatically per-formed by electronic sensors.
c. The high mean airway pressure alarm is the same on both models.
d. The inspiratory time control limits have been expanded on the 3100B.


ANS: B
The CareFusion 3100B has automatic piston-centering, and the center piston control has been eliminated from the 3100B. The 3100A has a mean pressure limit control; on the 3100B, this control has been eliminated. The high mean airway pressure alarm is 50 cm H2O on the 3100A and 60 cm H2O on the 3100B. The inspiratory time control limits are the same on both models.