The operational verification procedure (OVP) involves checking the ventilator circuit for leaks. Ventilator settings that could be used to perform this procedure include which of the following?

a. Tidal volume (VT) = 500 mL, Flow rate = 60 L/min, High pressure limit = 50 cm H2O
b. VT = 1000 mL, Flow rate = 20 L/min, High pressure limit = maximum
c. VT = 500 mL, Flow rate = 20 L/min, High pressure limit = maximum, Inspiratory pause = 2 seconds
d. VT = 200 mL, Flow rate = Maximum, High pressure limit = 50 cm H2O, Inspira-tory pause = 1 second


ANS: C
To check for leaks in the ventilator circuit the operator should set the tidal volume to 500 mL, the gas flow low (e.g., 20 L/min), the maximum pressure limit high (e.g., 100-120 cm H2O), and an inspiratory pause of 1-2 seconds.

Health Professions

You might also like to view...

What is the appropriate treatment regimen for a suspected tension pneumothorax?

A) Intubation, stabilization of the chest wall with dressings, manage shock B) Oxygen, IV fluids wide open, cardiac monitor, meds for dysrhythmias C) IV fluids to increase output, cardiac monitor, pericardiocentesis D) Ventilation, needle decompression, IV fluids to manage hypotension

Health Professions

Which of the following is NOT a factor affecting fire door construction and operation?

A. Type of occupancy B. Number of fire doors in the space C. Amount of space around door opening D. Required fire-protection rating for the door

Health Professions

Elongated cylinders that run up and down the long axis of the bone and have a central canal that contains blood vessels and nerves are called _______________. 

Fill in the blank(s) with the appropriate word(s).

Health Professions

The dividing point of a multirooted tooth is the ____________

a. grooves b. furcation c. fossa d. cingulum

Health Professions